LaMotte DC1100 Colorimeter Trouble Shooting Guide

Problem
Action
Check

Meter won't come on.

Battery?
  1. Turn meter over and remove four screws from the meter base.
  2. Remove old battery and insert new battery. The DC1100 uses a Type 1604A 9 volt Alkaline battery. NOTE: Must be alkaline battery
AC Adapter? To connect meter to a power source, plug the connector of the AC adapter into the small hole at the back of the meter. Plug large end into an electrical outlet (120V, 60Hz).
AC Wall Outlet? Verify power source.
Contact LaMotte for return authorization. Return to LaMotte for repair.

Comes on, but the readings drift

Are instrument grade reagents being used? Obtain instrument grade reagents.
Are there suspended solids in sample? Filter both blank and sample (after reaction). NOTE: Do not use filtered blank for sample reaction.
Do readings drift up scale? Bring samples to room temperature.
Readings drift down scale with cold sample? Check for sample tube fogging, let sample equilibrate to room temperature.

Suspect Calibration.

Will meter blank? (zero) If no, see below
Are instrument grade reagents being used? Obtain instrument grade reagents.
How old are the reagents? Contact LaMotte to verify age.
Check calibration with standards? How old is the standard?
Is class "A" volumetric glassware being used to make up the standards? If not, obtain and see Standard Methods.
Verify standard with alternate test method? Run alternate test on standard.
Verify with another meter? Check other meter calibrations.
Contact LaMotte for return authorization. Return for Calibration check.

Meter won't zero

Sample tubes free of dirt & scratches? Check & replace if necessary.
Sample tubes clean? Clean tubes often with soap and a test tube brush. Do not leave reacted samples in tubes, staining may occur.
Possible internal meter components subjected to moisture? Always dry tubees before inserting. Examine tube chamber for visible moisture or chemical stains. If yes, remove 4 screws from back, split case & dry with hair dryer under low heat (10-15 min).
LED contaminated with dirt and deposits? Clean square LED and Photodetector w/cotton swab & small amount of window cleaner. EXAMPLE: Windex®
High suspended solids in sample blank? Filter with coffee filter, or similar, to remove most solids. (Be sure to filter reacted samples as well) NOTE: Do not use filtered blank for sample reaction.
At left of display meter reads 1-w/o samples in chamber Meter over range? Contact LaMotte for return authorization & return.
At left of display meter reads 1-only with samples in chamber Meter over range? Extremely high concentration, dilute sample to see if 1- condition returns to normal condition.